Detection of hypoglycin A and MCPA‐carnitine in equine serum and muscle tissue: optimisation and validation of a LC‐MS based method without derivatisation

González Medina, S; Hyde, C; Lovera, I; Piercy, R J

Authors

S González Medina

C Hyde

I Lovera

R J Piercy



Abstract

Measurement of hypoglycin A (HGA) and its toxic metabolite, methylenecyclopropylacetic acid (MCPA), in equine serum confirms a diagnosis of atypical myopathy (AM), a pasture‐associated toxic rhabdomyolysis with high mortality linked to the ingestion of Acer trees plant material. Supportive diagnostic tests include plasma acyl‐carnitine profiling and urine organic acid testing, but these are not specific for AM. Previously reported HGA and MCPA analytical techniques used liquid chromatography–mass spectrometry (LC‐MS) with a derivatising step, but the latter prolongs testing and increases costs.

Objectives
To develop a rapid LCMS method for detection of serum and tissue HGA and MCPA that enables expedited diagnosis for horses with AM.

Study design
Analytical test validation.

Methods
Validation parameters to industry standards using as criteria precision, accuracy, linearity, reproducibility and stability in analyte‐spiked samples were calculated on 9‐calibration points and 3 different validation concentrations in both serum and muscle tissue.

Results
The test was successfully validated for the detection of HGA and MCPA‐carnitine in horse serum and muscle. Test linearity was excellent (r2=0.999), accuracy was very good for both analytes (93‐108%), precision did not exceed 10% coefficient of variation and reproducibility met the requirements of the Horwitz equation. Stability was unaffected by storage at a range of temperatures.

Main limitations
The spectrum of the tested analytes was limited to only two relevant analytes in favour of a quick and easy analysis. Linearity of the muscle method was not evaluated as calibration curves were not produced in this matrix.

Conclusion
We report an optimised, simplified and validated method for detection of HGA and MCPA‐carnitine in horse serum and muscle suitable for rapid diagnosis of suspected AM cases. The serum based test should also enable risk assessment of toxin exposure in co‐grazing horses and assessment of horses with undiagnosed myopathies, while the tissue detection test should help to confirm cases post‐mortem and to determine toxin distribution, metabolism and clearance across different tissues.

Citation

González Medina, S., Hyde, C., Lovera, I., & Piercy, R. J. (2020). Detection of hypoglycin A and MCPA‐carnitine in equine serum and muscle tissue: optimisation and validation of a LC‐MS based method without derivatisation. Equine Veterinary Journal, https://doi.org/10.1111/evj.13303

Journal Article Type Article
Acceptance Date May 25, 2020
Publication Date Jun 11, 2020
Deposit Date Aug 14, 2020
Publicly Available Date Aug 14, 2020
Journal Equine Veterinary Journal
Print ISSN 0425-1644
Publisher Wiley
Peer Reviewed Peer Reviewed
DOI https://doi.org/10.1111/evj.13303
Public URL https://rvc-repository.worktribe.com/output/1377084
Publisher URL https://doi.org/10.1111/evj.13303

Files





You might also like



Downloadable Citations